- LCD display of 128×64 matrix is used.
- Using the D high precision impact
device
- Converts to all common hardness scales(HV
,HB,HRC,HRB,HRA,HS).
- English displaying and menu operating, the operation
is easy and convenient.
- Hermetically sealed metal
shell, solid and durable, has excellent resistance to vibration, shock and
electromagnetic interference ability.
- Equipped with a data
processing software, through the USB interface, can exchange data with computer
printing and parameter setting
- Upper and lower limit of hardness can be preset; When the
tested value exceeds the limits, alarm will send out automatically to make
convenient for the requirements of batch measurements. Back light display has
been used to make convenient for the use in poor light.
- Material of “cast steel” is added; HB values can be read
out directly when D/DC impact device is used to measure “cast steel” work
piece.
- Printer be separated from main unit and copies of testing
results can be printed as required.
- Power is two of AA battery. Continuous working period:
approx. 200 h (no back light on).
- Software of PC can be installed according to the
requirements of user, the function will be more powerful to satisfy the more
strict demands of quality control and management.
- Failure analysis of pressure vessel, steam
turbo-generator set and other equipment
- With USB2.0 interface, multiple communication modes
are adopted to meet customized requirements of various users.
- Equipped with 7 types of impact devices which need
not to be recalibrated when changing them, the system can identify the
type of impact device automatically.
- Max 600 groups(impact times:32~1)of data can be stored at internal non volatile data
storage.
- Test values software calibration function.
- Narrow
testing space where work piece installed
Applicable materials:
Steel and cast steel, alloy tool steel, stainless steel,
gray cast iron, nodular cast iron, cast aluminum alloy, copper zinc alloys
(brass), an alloy of copper and tin, copper (bronze), forged steel.
Main
Application:
- The assembled machinery and permanently installed
parts
- Die cavity of molds
- Heavy work piece
- Bearings and other parts
- Cases which require the test result with normalized
original recording
- Material identification of the metal material
warehouse
- Quick tests of large range and multipoint measuring
positions for heavy Workpiece.
The KH200 portable Leeb hardness tester-Features:
- Using the D high precision
impact device
- Hermetically sealed metal
shell, solid and durable, has excellent resistance to vibration, shock and
electromagnetic interference ability.
- Shaw (HS), a (HL), (HB)
Brinell, Rockwell (HRC/HRB/HRA), Vickers (HV) and other 7 kinds of
hardness value realization of a measurement, automatic conversion
- Has a starting value
software calibration function, ensure the data in any industry codes and
standards applicable and accurate.
- The optional 7 different
impact device, when the replacement does not require calibration,
automatic recognition
- Can be pre-set limit value,
overrun alarm
- Can store 600 groups of
measured value of all Chinese menu, backlit display, convenient operation
- Equipped with a data
processing software, through the USB interface, can exchange data with
computer printing and parameter setting
Technical parameters:
Measuring
Range
|
HLD(170-960)
(170-960)HLD,(19-651)HB,(13-100)HRB,(17-68.5)HRC,(59-85)HRA,(80-976)HV、(30~110)HS
|
Impact
Directions
|
up, down, horizontal,
slanting up, slanting down
|
Measuring
materials
|
Steel and cast steel,
alloy tool steel, stainless steel, gray cast iron, nodular cast iron, cast
aluminum alloy, copper zinc alloys (brass), an alloy of copper and tin,
copper (bronze), forged steel
|
Hardness
scales
|
HL, HB, HRB,
HRC, HRA, HV,HS.
|
Display
|
128*64
digital matrix LCD
|
Memory
|
600group(32
every group)
|
Work voltage
|
3V( 2 AA
alkaline batteries)
|
Interface
port
|
USB 2.0
|
Dimensions
|
125mm*71mm*27mm
|
Weight
|
0.3KG
|
Continuous
working period
|
200 hours
(with backlight off)
|
Accuracy and
repeatability of displayed value:
No.
|
Type of
impact device
|
Hardness
value of Leeb standard hardness block
|
Error of displayed
value
|
Repeatability
of displayed value
|
1
|
D
|
760±30HLD
530±40HLD
|
±6 HLD
±10 HLD
|
6HLD
10HLD
|
2
|
DC
|
760±30HLD
530±40HLD
|
±6 HLDC
±10 HLDC
|
6HLD
10HLD
|
3
|
DL
|
878±30HLD
736±40HLD
|
±12 HLDL
|
12HLDL
|
4
|
D+15
|
766±30HLD
544±40HLD
|
±12 HLD+15
|
12HLD+15
|
5
|
G
|
590±40HLD
500±40HLD
|
±12 HLG
|
12HLG
|
6
|
E
|
725±30HLD
508±40HLD
|
±12 HLE
|
12HLE
|
7
|
C
|
822±30HLD
590±40HLD
|
±12 HLC
|
12HLC
|
Standard
configuration:
|
No.
|
Item
|
Quantity
|
Standard
Configuration
|
1
|
Main
body
|
1
|
2
|
D
type impact device
|
1
|
3
|
Test block with HLD value
|
1
|
4
|
Cleaning
brush (I)
|
1
|
5
|
Instruction manual
|
1
|
6
|
Warranty card
|
1
|
7
|
Small
supporting ring
|
1
|
8
|
Screwdriver
|
1
|
9
|
Certificate of conformity
|
1
|
10
|
Instrument box
|
1
|
11
|
5 alkaline battery
|
2
|
Optional
Configuration
|
1
|
Other types of impact
devices
|
7
|
2
|
Other types of support
ring
|
12
|
3
|
Other types of impact
ball/body
|
2
|
4
|
Data processing software
|
1
|
5
|
Connecting line
|
1
|
|